Precision in programmed infusion flow of enteral nutrition pumps

Authors

  • Juan Ignacio Padilla Cuadra Caja Costarricense del Seguro Social, Hospital Dr. Rafael A. Calderón Guardia
  • María Luisa Fallas Quirós Caja Costarricense del Seguro Social, Hospital Dr. Rafael A. Calderón Guardia
  • Cecilia Vargas Hernández Caja Costarricense del Seguro Social, Hospital Dr. Rafael A. Calderón Guardia
  • Jorge Ramírez Arce Caja Costarricense del Seguro Social, Hospital Dr. Rafael A. Calderón Guardia
  • Marieta Arias Benavides Caja Costarricense del Seguro Social, Hospital Dr. Rafael A. Calderón Guardia

DOI:

https://doi.org/10.51481/amc.v50i3.382

Keywords:

infusion pumps , enteral nutrition

Abstract

Justification and aim: To evaluate in vitro the accuracy of volume infusion of peristaltic enteral nutrition pumps to determine if this variable may be responsible for error's in the administration of nutrition support.
Materials and Methods: Thirteen pumps were evaluated through simulation using a model that resembles the administration of enteral nutrition to a patient. For every case, we used an standard formula diluted as recommended by the manufacturer. A volume of 240 mL was programmed to be infused at 40 ml per hour. For each case, time was measured until finishing the volume, comparing to the programmed time and calculating the delivery error.
Results: In 12 of 13 pumps, we detected error by excess, which varied from 7.5% to 52.5%. Only one pump showed an error of 0%. None of the pumps demonstrated infusion less than programmed.

Conclusion: Most of the pumps analysed demonstrate high percentage of error in comparison to what is accepted by the manufacturer.

Downloads

Download data is not yet available.

References

Jolliet P, Pichard C, Biolo G, Chiolero R, Grimble G, Leverve X et al . Enteral nutrition in intensive care patients: a practical approach. Working group on nutrition and metabolism, ESICM. European Society of Intensive Care Medicine. Intensive Care Med 1998; 24: 848-859.

Barr J, Hecht M, Flavin K, Khorana A, Gould M. Outcomes in critically ill patients before and after the implementation of an evidence-based nutritional management protocol. Chest 2004; 125: 1446-1457.

Heyland DK, Dhaliwal R, Drover JW, Gramlich L, Doder P, Canadian critical care clinical practice guidelines committee: canadian clinical practice guidelines for nutrition support in mechanical ventilated, critically ill adult patients. J Parenter Enteral Nutr 2003: 27: 355-373.

Adam S, Baston S. A study of problems associated with the delivery of enteral feed in critically ill patients in five ICU´s in the UK. Intensive Care Med. 1997: 23: 261-266.

Santana L, Oshanahan O, Garcíal M, Ramírez A, Sánchez M, Hernández E. Calidad del soporte nutricional artificial en una unidad de cuidados intensivos. Nutr Hosp 2006; 21: 661-666.

Garvin CG, Brown R.O. Nutritional support in the intensive care unit: are patients receiving what is prescribed ? Crit Care Med 2001; 20:204-205.

McClave SA, Sexton LK, Spain DA, Adams JL, Owens NA, Sullins MB, et al. Enteral tube feeding in the intensive care unit: factor impending adequate delivery. Crit Care Med 1999; 27: 1252-1256.

De Jonghe B, Appere-De-Verdi C, Fournier M, Tran B, Mener J, Melchior JC, Outin H. A prospective survey of nutritional support practices in intensive care unit patients: what is prescribed ? What is delivered?. Crit Care Med 2001; 29: 8-12.

Binnekade JM, Tepaske R, Brynzeld P, Mathus Vliegen EMH, Hann RJ. Daily enteral feeding on the ICU: attainment of goals ant interfering factors. Crit Care Med 2005: 9: 218-225.

De B, Chapman M, Fraser R, Finnie M, De Keulenaer B, Liberalli D, Satanek M: Enteral nutrition in the critically ill: a prospective survey in Australian intensive care unit. Anesth Intensive Care 2001; 29: 619-622.

Heyland DK, Schroter-Noppe D, Drover JW, Jain M, Keefe L,

Dhalival R, et al. Nutrition support in the critical care setting: Current practice in Canadian ICU´s - Opportunities for improvement ? J Parent Enter Nutr 2003; 27: 74-83.

Ciocon J, Galindo-Ciocon D, Thiessen C, Comparison of intermittent versus continuous tube feeding among the elderly. J Parent Enter Nutr 1992; 16: 525-528.

Jones BJ, Payne F. Silk DB. Indications for pumps assisted enteral feeding. Lancet 1980; 1: 1057-1058.

Guenter P, Silkorski M, Tube feeding. Practical guidelines and nursing protocols. ASPEN Publishers Inc . Gaitherburg, Maryland 2001.

Tepaske R, Binnekade J, Goedhart P, Schulz M, Vroom M, Mathus Vliegen E MH. Clinically relevant differences in accuracy of enteral nutrition feeding pump systems. J Parent Enter Nutr 2006; 30 : 339343.

Downloads

Published

2008-07-01

How to Cite

Precision in programmed infusion flow of enteral nutrition pumps. (2008). Acta Médica Costarricense , 50(3), 168-171. https://doi.org/10.51481/amc.v50i3.382