Experience with special pediatric contrast radiological studies and comparison of operational indicators

Authors

DOI:

https://doi.org/10.51481/amc.v67i3.1469

Keywords:

fluoroscopy, radiology, pediatric, radiological protection, process optimization

Abstract

Objective:  With the increased use of imaging techniques in pediatric care, a population considered more radiosensitive than adults, the optimization of contrast-enhanced radiological studies is essential. The objective was to describe the experience with specialized pediatric contrast-enhanced radiological studies performed at the Radiology Department of the Dr. Carlos Sáenz Herrera National Children's Hospital and to compare clinical practice with international benchmarks.

Methods:  This was an observational, cross-sectional study, recording all procedures performed, including their technical and procedural parameters, from March 15 to June 15, 2023. Information was collected on the types of procedures, the number of images generated per study, patient age, and fluoroscopy time. Fluoroscopy time and number of images generated per study were defined based on the distribution of the median values ​​for fluoroscopy time and number of images. The tabulated data, classified by study type and age range, were compared with values ​​reported by international references, obtained through a targeted literature review. A separate literature review of international and national theoretical frameworks on radiation protection was also conducted to develop a proposal for technical optimization.

Results: A total of 511 pediatric special contrast radiological studies were included: 167 voiding cystourethrograms, 126 gastrointestinal tract studies, 124 esophagograms and 94 contrast enemas.  With serial voiding cystourethrography, an average of 35.8 images were recorded, with an average fluoroscopy time of 17 seconds and an exposure time ranging from 2 to 195 seconds, with a median of 12 seconds. For the other barium contrast studies, the average number of images acquired was 93.6, with an average fluoroscopy time of 31.2 seconds, ranging from 1 to 229 seconds; specifically, the median was 22 seconds for the barium enema and 20.5 seconds for the esophagogram. Comparisons with all reference studies consistently showed shorter fluoroscopy times and a higher number of images generated per study in Costa Rica. A series of 14 actions was compiled to guide protection strategies.

 Conclusions: Fluoroscopy times used in the country are shorter than those typically employed in other contexts for the selected studies; however, the number of intraprocedural radiographs taken locally was higher. As part of the strategy to improve radiation protection, it is suggested that fluoroscopy times be slightly increased to reduce the number of radiographs, thereby optimizing the balance between diagnostic quality and radiation exposure.

Downloads

Download data is not yet available.

References

1. International Commission on Radiological Protection (ICRP). Avoidance of radiation injuries from medical interventional procedures. ICRP publication 85. Ann ICRP. 2000;30: 1–66.

2. International Commission on Radiological Protection (ICRP). Radiological protection in paediatric diagnostic and interventional radiology. ICRP Publication 121. Ann ICRP. 2013;42: 1–63.

3. International Commission on Radiological Protection (ICRP). Diagnostic reference levels in medical imaging. ICRP Publication 135. Ann ICRP. 2017;46:1–144.

4. European Society of Radiology/European Commission Radiation Protection. European guidelines on diagnostic reference levels for paediatric imaging. No. 185. Luxembourg: Publications Office of the European Union; 2018. DOI:10.2833/003998.

5. Gomes Y, Brown G, Davies J, Parange N, Chau M. Establishing diagnostic reference levels for pediatric fluoroscopic examinations in a tertiary hospital. Pediatr Radiol. 2022;52: 1296–1305. DOI:10.1007/s00247-022-05290-1.

6. Chytrá K, Novák L, Rejtar P, Homola M, Daní?ková K, Cervinková I. Radiation exposure of paediatric general radiography, fluoroscopy and CT procedures in the Czech Republic-pilot study. Radiat Prot Dosimetry. 2019;1–6. DOI:10.1093/raddos/ncz238.

7. Ashworth E, Woods L, Cook V. Diagnostic reference levels in paediatric fluoroscopy: how does a secondary referral centre compare with 2018 European guidelines. Br J Radiol. 2021; 94: 20201269. DOI: 10.1259/bjr.20201269.

8. Sulieman A, Alzimami K, Elhag B, Babikir E, Alsafi K. Evaluation of radiation dose to pediatric patients during certain special procedures. Radiat Phys Chem. 2014;104:267–271. DOI: 10.1016/j.radphyschem.2013.11.034.

9. Bibbo G, Balman D, Linke R. Diagnostic reference levels for common paediatric fluoroscopic examinations performed at a dedicated paediatric Australian hospital. J Med Imaging Radiat Oncol. 2016;60: 469–474. DOI: 10.1111/1754-9485.12478.

10. Hernanz-Schulman M, Goske MJ, Bercha IH, Strauss KJ. Pause and pulse: Ten steps that help manage radiation dose during pediatric fluoroscopy. Am J Roentgenol. 2011;197: 475–481. DOI: 10.2214/AJR.10.6122.

11. Lazarus MS, Taragin BH,Malouf W, Levin TL, Nororis E, Schoenfeld AH, et al. Radiation dose monitoring in pediatric fluoroscopy: comparison of fluoroscopy time and dose–area product thresholds for identifying high-exposure cases. Pediatr Radiol. 2019;49: 600–608. DOI: 10.1007/s00247-018-04335-8.

12. Sulieman A, Elhag B, Alkhorayef M, Babikir E, Theodorou K, Kappas C, et al. Estimation of effective dose and radiation risk in pediatric barium studies procedures. Appl Radiat Isot. 2017;138: 40-44. DOI: 10.1016/j.apradiso.2017.07.013.

13. Livingstone RS, Eapen A, Chiramel GK. Radiation dose to paediatric patients undergoing fluoroscopic examinations performed using digital imaging system. Radiography. 2008;14: 17–23. DOI: 10.1016/j.radi.2006.06.004.

14. Tzanis E, Raissaki M, Konstantinos A, Damilakis J, Perisinakis K. Radiation exposure to infants undergoing voiding cystourethrography: the importance of the digital imaging technology. Phys Med. 2021;85:123–128. DOI: 10.1016/j.ejmp.2021.05.006.

15. Ghodadra A, Bartoletti S. Reducing radiation dose in pediatric diagnostic fluoroscopy. J Am Coll Radiol. 2016;13:55–58. DOI: 10.1016/j.jacr.2015.07.03.

16. Hayashi S, Takenaka M, Hosono M, Kogure H, Hasatani K, Suda T, et al. Diagnostic Reference Levels for Fluoroscopy-guided Gastrointestinal Procedures in Japan from the REX-GI Study: A Nationwide Multicentre Prospective Observational Study. Lancet Reg Health West Pac. 2022;20: 100376. DOI: 10.1016/j.lanwpc.2021.100376.

17. Vañó E, Miller DL, Martin CJ, Rehani MM, Kang K, Rosenstein M, et al. ICRP Publication 135: Diagnostic Reference Levels in Medical Imaging. Ann ICRP. 2017;46: 1-144. DOI: 10.1177/0146645317717209.

18. Fisher RF, Applegate KE, Berkowitz LK, Christianson O, Dave JK, DeWeese L, et al. AAPM Medical Physics Practice Guideline 12.a: Fluoroscopy Dose Management. J Appl Clin Med Phys. 2022;23:e13526. DOI: 10.1002/acm2.13526.

Published

2026-03-01

How to Cite

Experience with special pediatric contrast radiological studies and comparison of operational indicators. (2026). Acta Médica Costarricense , 67(3), 1-10. https://doi.org/10.51481/amc.v67i3.1469